With ACQUITY Solvent Manager, pump fails dynamic leak test - WKB231485

Article number: 231485

SYMPTOMS

  • Accumulator pump fails dynamic leak test
  • Primary Pump fails DLT, Accumulator passes consistently
  • Random pressure drops during flow
  • High-pressure seals and plunger are new
  • Head and Seal wash housing are new

ENVIRONMENT

  • ACQUITY UPLC Binary Solvent Manager (BSM)
  • ACQUITY H-Class Quaternary Solvent Manager (QSM)
  • ACQUITY UPLC H-Class BSM
  • ACQUITY UPLC I-Class BSM

CAUSE

Faulty accumulator check valve

FIX or WORKAROUND

  1. Replace the accumulator check valve.
  2. Prime the solvent line and run some flow to confirm stable system pressure and low pressure delta.
  3. Perform the dynamic leak test to ensure that the leak rate is within acceptable limits.

ADDITIONAL INFORMATION

ACQUITY UPLC BSM Accumulator check valve part number: 700002968

ACQUITY UPLC I-Class BSM Accumulator check valve part number: 700003755

ACQUITY UPLC H-Class BSM Accumulator check valve part number: 700003755

ACQUITY UPLC H-Class QSM Accumulator check valve part number: 700005164

ACQUITY UPLC H-Class Bio QSM Accumulator check valve part number: 700005415

 