With ACQUITY Solvent Manager, pump fails dynamic leak test - WKB231485
Article number: 231485
SYMPTOMS
- Accumulator pump fails dynamic leak test
- Primary Pump fails DLT, Accumulator passes consistently
- Random pressure drops during flow
- High-pressure seals and plunger are new
- Head and Seal wash housing are new
ENVIRONMENT
- ACQUITY UPLC Binary Solvent Manager (BSM)
- ACQUITY H-Class Quaternary Solvent Manager (QSM)
- ACQUITY UPLC H-Class BSM
- ACQUITY UPLC I-Class BSM
CAUSE
Faulty accumulator check valve
FIX or WORKAROUND
- Replace the accumulator check valve.
- Prime the solvent line and run some flow to confirm stable system pressure and low pressure delta.
- Perform the dynamic leak test to ensure that the leak rate is within acceptable limits.
ADDITIONAL INFORMATION
ACQUITY UPLC BSM Accumulator check valve part number: 700002968
ACQUITY UPLC I-Class BSM Accumulator check valve part number: 700003755
ACQUITY UPLC H-Class BSM Accumulator check valve part number: 700003755
ACQUITY UPLC H-Class QSM Accumulator check valve part number: 700005164
ACQUITY UPLC H-Class Bio QSM Accumulator check valve part number: 700005415