Instrument power lights do not come on - WKB230665
Article number: 230665
SYMPTOMS
- Instrument power lights do not come on
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY H-Class PLUS
- ACQUITY I-Class PLUS
- ACQUITY M-Class
- Alliance
- Alliance iS HPLC System
- Xevo
- Synapt
- ACQUITY QDa/QDa II
- ASAP RADIAN
- ACQUITY UPLC Tunable UV Detector
- ACQUITY UPLC PDA Detector
- ACQUITY UPLC PDA eλ Detector
- ACQUITY UPLC FLR Detector
- ACQUITY UPLC ELS Detector
CAUSE
Inadequate lab power where the instruments are plugged into the wall circuit or power strip
FIX or WORKAROUND
- Swap out the surge protector or battery backup and plug into the wall directly, or use an extension cord and plug into another circuit known to be working to test.
- Have a licensed electrician check the power output from the problem outlet.