ACQUITY UPLC Solvent Manager passes dynamic leak test with negative leak rate on primary pump - WKB230615

Article number: 230615

SYMPTOMS

  • Successful dynamic leak test on accumulator
  • Dynamic leak test "passes" on primary pump with zero and/or negative leak rate values

ENVIRONMENT

  • ACQUITY UPLC Binary Solvent Manager (BSM)
  • ACQUITY UPLC I-Class BSM

CAUSE

Faulty i2V cartridge on primary pump

FIX or WORKAROUND

  1. Replace the i2V cartridge (1/pk, 700005165; 2/pk 700003563).
  2. Perform the dynamic leak test to confirm functionality.