Leak at check valve of 1525 Binary HPLC Pump - WKB230565

Article number: 230565

SYMPTOMS

  • Pump B leaks at check valve and tubing
  • Tightening fitting does not resolve                                                                                                                                                                   clipboard_ea2e959b4d6140b27bd9c0ede6ec6c26b.png                                                                                                                                                 clipboard_ef0617b5677470f404312b3f0ebe6d923.png

ENVIRONMENT

  • 1525 Binary HPLC Pump

CAUSE

Worn fittings

FIX or WORKAROUND

  1. Replace the check valve housing: 
    • 700001108 - check valve housing for regular heads
    • WAT025216 - check valve housing for extended flow heads
  2. Replace tubing, outlet LH (430001254).