SM-FTN leaks from port 5 on injection valve cartridge - WKB229847

Article number: 229847

SYMPTOMS

  • SM-FTN leaks from port 5 on injection valve cartridge
  • Visible leak even at low flow rates

ENVIRONMENT

  • ACQUITY H-Class
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC Sample Manager - Flow Through Needle 
  • ACQUITY UPLC H-Class PLUS Bio

CAUSE

The pump outlet tubing fitting into port 5 on the injection valve cartridge is damaged.

FIX or WORKAROUND

  1. Replace the pump outlet tubing (part 430002473 for H-Class, 430002901 for I-Class).
  2. Replace the injection pod if the connection is worn out (part 700006057 for I-Class, 700005236 for H-Class).

ADDITIONAL INFORMATION

Put flow through the system and check for leaks.