Alliance 2695 shows many chromatogram spikes - WKB228408
Article number: 228408
SYMPTOMS
- Consecutive spikes in chromatogram baseline
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- ARC HPLC
- ACQUITY ARC
CAUSE
Flow path/solvent lines dirty from from usage (different solvent and/or samples)
FIX or WORKAROUND
- Prepare Magic Mix (25% each of H2O; MeOH, ACN, IPA) and add up to 10% formic acid.
- Remove the column from the system.
- Prime the solvent manager and the needle wash with Magic Mix.
- Apply a flow of 0.5 mL/min and leave Magic Mix on the system overnight.
- Prime all lines to 100% H2O.
- Apply a flow of 0.5 mL/min for 30 minutes.
- Prepare new solvents and set up your system as needed.
- Check the baseline.
ADDITIONAL INFORMATION