Alliance 2695 shows many chromatogram spikes - WKB228408

Article number: 228408

SYMPTOMS

  • Consecutive spikes in chromatogram baseline

clipboard_e3de4d848c49b4feba0ecd6c19576b937.png

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • ARC HPLC
  • ACQUITY ARC

CAUSE

Flow path/solvent lines dirty from from usage (different solvent and/or samples)

FIX or WORKAROUND

  1. Prepare Magic Mix (25% each of H2O; MeOH, ACN, IPA) and add up to 10% formic acid.
  2. Remove the column from the system.
  3. Prime the solvent manager and the needle wash with Magic Mix.
  4. Apply a flow of 0.5 mL/min and leave Magic Mix on the system overnight.
  5. Prime all lines to 100% H2O.
  6. Apply a flow of 0.5 mL/min for 30 minutes.
  7. Prepare new solvents and set up your system as needed.
  8. Check the baseline.

ADDITIONAL INFORMATION

 