Retention time shift with ACQUITY H-Class QSM - WKB22815
Article number: 22815
SYMPTOMS
-
Retention time shift
-
Dynamic leak test failure when end point is inject pod outlet
- Strange pressure profiles across gradient
ENVIRONMENT
-
ACQUITY UPLC H-Class QSM
-
ACQUITY UPLC H-Class QSM Bio
CAUSE
Restriction in mixer and/or the tubing from vent valve to mixer
FIX or WORKAROUND
Replace the mixer.
ADDITIONAL INFORMATION
700005177 - ASSY, MIXER, 100µL, QSM (Default)
205000731 - ASSY, MIXER, 22µL, QSM
205000719 - ASSY, MIXER, 250µL, QSM
700005258 - ASSY, MIXER, 100µL, bioQSM