FTN needle hangs and lifts the vials/tray - WKB228101

Article number: 228101

SYMPTOMS

  • Missing vials error when injecting
  • Sample vials are lifted out of the tray and scattered in the sample manager

ENVIRONMENT

  • ACQUITY APC Sample Manager - FTN
  • ACQUITY Arc Sample Manager FTN-R
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS

CAUSE

Needle was not routed correctly. It hangs too low and pulls vials out of the tray.

FIX or WORKAROUND

Re-route the needle properly, or replace the needle if it is too badly affected.

ADDITIONAL INFORMATION