String error and SM error in console - WKB225989
Article number: 225989
SYMPTOMS
- String error and then SM error appear in console following reset
- Column temperature is missing from the SM console page
ENVIRONMENT
- Classic ACQUITY
- Empower
CAUSE
A shortcut was used to open the Console before opening Empower. If the console is opened from the Run Samples window in Empower, no fault is seen and the Console opens and works correctly. When the console is opened before Empower using the shortcut, errors occur.
FIX or WORKAROUND
Remove the pinned shortcut from the taskbar and open the console through Empower.
ADDITIONAL INFORMATION
The shortcut previously worked OK, but for some reason started to fail.