High back pressure on all column paths - WKB223996
Article number: 223996
SYMPTOMS
- High back pressure on all column paths
- High back pressure without column
- Tubing after the column can pop out
ENVIRONMENT
- ACQUITY UPLC Column Manager
CAUSE
The tubing after the second CM valve to the detector inlet is blocked.
FIX or WORKAROUND
- Try cutting both ends of the tubing and backflushing. If it works, you're done.
- Replace the tubing (700003443 - Tubing, .062 x 0.0025, PEEK, 60").
ADDITIONAL INFORMATION
700003443 - Tubing, .062 x 0.0025, PEEK, 60"
700003799 - Tubing, .062 x .004, PEEK, 60" (Standard)
