Cannot shut off the lamp using 2998 method events in Empower (Japanese version only) - WKB224053

Article number: 224053

SYMPTOMS

  • Cannot shut off lamp using method events

ENVIRONMENT

  • Issue is reproducible in ICS 1.73.2643, 2998 firmware 1.70.6363 (DP 2019 - DP 2021)
  • Empower 3 FR4 Japanese
  • Empower 3 FR5 SR4 Japanese

CAUSE

This is a translation issue in the driver. CRI-4141 was created and accepted in the backlog for development engineering. There is no planned timeline for the fix.

FIX or WORKAROUND

  1. Create a separate instrument method and method set with the lamp turned off. 
  2. Add this method set at the end of the sample set.

ADDITIONAL INFORMATION

 