Skip to main content
Waters

Sample vial is lifted from plate position during injection on ACQUITY UPLC - WKB223935

Article number: 223935

SYMPTOMS

  • During an injection, the sample vials are lifted from their plate positions by the needle
  • "Z-axis h/w lost steps" or "Z-axis move h/w fault" error

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC Sample Manager 
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

Bent needle

FIX or WORKAROUND

Replace the needle.

ADDITIONAL INFORMATION

Standard ACQUITY Sample Manager Needles

  • 700002644 - 30 microL PEEK: Assy, Needle, Sample, .010 I.D.
  • 205000362 - SS: Sample Needle Kit, Stainless Steel (30 µL)

Standard ACQUITY H-Class FTN Needles

  • 700005215 - Needle, 15µL SS w/ Guide
  • 700011628 - PLUS: Sample Needle Kit, 15µL Textured
  • 700005421 - Bio: Needle, 15µL MP35N w/ Guide
  • 700011658 - Bio PLUS: Sample Needle Kit, 15µL Textured (Bio)

Standard ACQUITY I-Class FL Needles

  • 700005923 - Kit, I-Class PEEK Needle 10µL
  • 700005924 - Kit, I-Class Steel Needle, 10µL

Standard ACQUITY I-Class FTN Needles

  • 700008977 - Sample Needle Kit, 15µL, SS
  • 700011628 - PLUS: Sample Needle Kit, 15µL Textured

Not able to find a solution? Click here to request help.