Alliance 2695 fails area and height precision with 25-µL syringe - WKB223819
Article number: 223819
SYMPTOMS
- 2695 fails height and area precision with 25-µL syringe
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2695D Dissolution Separations Module
CAUSE
Suction of solution from the vial is too fast.
FIX or WORKAROUND
In the Instrument Method Editor, on the General tab, select the Slow option for the syringe draw rate (µL/sec) parameter.