Shift in retention time on ACQUITY UPLC Sample Manager - Flow Through Needle PLUS - WKB222998
Article number: 222998
SYMPTOMS
- Shift in retention time on ACQUITY UPLC Sample Manager - Flow Through Needle PLUS
- Solvent Manager Dynamic leak tests pass
- Needle Seal Readiness leak test fails
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
The extension loop connecting to the sample needle loose
FIX or WORKAROUND
- Tighten up extension loop fittings
- Replace needle and or extension loop if fittings are damaged or worn
ADDITIONAL INFORMATION