Shift in retention time on ACQUITY UPLC Sample Manager - Flow Through Needle PLUS - WKB222998

Article number: 222998

SYMPTOMS

  • Shift in retention time on ACQUITY UPLC Sample Manager - Flow Through Needle PLUS 
  • Solvent Manager Dynamic leak tests pass
  • Needle Seal Readiness leak test fails

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

The extension loop connecting to the sample needle loose

FIX or WORKAROUND

  1. Tighten up extension loop fittings 
  2. Replace needle and or extension loop if fittings are damaged or worn

ADDITIONAL INFORMATION

 

id222998, A-30SM, DLT, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPPFTN, UPSMFTN, UPSMFTN+