Gradient does not change or start on ACQUITY M-Class - WKB223056

Article number: 223056

SYMPTOMS

  • Installation - so it never worked
  • Start acquisition and the gradient never changes from the initial gradient

ENVIRONMENT

CAUSE

Events and Triggering setup 

FIX or WORKAROUND

  1. Do NOT run an events and triggering contact closure cable between the M-Class and the MS.
  2. In the Inlet Editor, select Instrument Configuration from the Tools drop-down menu.
  3. Select Events & Triggering.
  4. Select Next.
  5. Select Events In 1.
  6. Select "Trigger by Contact Closure" under Pump.
  7. Select "Trigger by Software" under Detector. 
  8. Select Next.
  9. Select Finish > Finish.
  10. Run samples.

ADDITIONAL INFORMATION

 

id223056, SCN1009, SCN1015, SYNAPTXS, UPMASM, UPMBSM, UPMSMFL