Gradient does not change or start on ACQUITY M-Class - WKB223056
Article number: 223056
SYMPTOMS
- Installation - so it never worked
- Start acquisition and the gradient never changes from the initial gradient
ENVIRONMENT
- ACQUITY M-Class
- SYNAPT XS
- 2019 Driver Pack Release 2 Service Release 1 Microscale
- MassLynx 4.2 (SCN 1009 or SCN 1015)
CAUSE
Events and Triggering setup
FIX or WORKAROUND
- Do NOT run an events and triggering contact closure cable between the M-Class and the MS.
- In the Inlet Editor, select Instrument Configuration from the Tools drop-down menu.
- Select Events & Triggering.
- Select Next.
- Select Events In 1.
- Select "Trigger by Contact Closure" under Pump.
- Select "Trigger by Software" under Detector.
- Select Next.
- Select Finish > Finish.
- Run samples.
ADDITIONAL INFORMATION
id223056, SCN1009, SCN1015, SYNAPTXS, UPMASM, UPMBSM, UPMSMFL