3767 SM does not prime system - WKB220311
Article number: 220311
SYMPTOMS
- 3767 does not prime system
- Default inlet method configured as expected with good communication to all modules
- 3767 diagnostics function as expected
ENVIRONMENT
- 3767 sample manager
- MassLynx SCN1036
CAUSE
Corrupted inlet method
FIX or WORKAROUND
Create a new inlet method and save.
ADDITIONAL INFORMATION
id220311, 3767, comms, SCN 1036