How much DMSO can I use? - WKB219633
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC Tunable UV Detector
- ACQUITY UPLC PDA Detector
- 2695 Separations Module
- 2695e Separations Module
ANSWER
DMSO is a strong solvent that is able to solubilize a lot of different analytes. Typically, using DMSO in small amounts is okay. However, if used in larger amounts, it can cause non-metallic tubings, especially PEEK, to swell.
Can I use DMSO as a diluent?
Yes, using DMSO as a diluent is okay as this is only a small amount. - WKB14943
DMSO is also commonly injected to clean columns. Make sure the column is compatible with DMSO. - WKB65747
PEEK needles should be changed to metallic ones. - WKB28265
If the needle is PEEK, keep the diluent to less than 10%.
Can I use DMSO in solvent?
Without the THF/hexane kit, limit DMSO to less than 10%. It is acceptable to use DMSO as an additive in low amounts.
With THF/hexane kit, DMSO can be used. Although more frequent PMs may be required.
Will the flow cell be affected if DMSO is used?
It depends on how much is being used.
Most flow cells contain PEEK tubing, so it is always best to limit DMSO to less than 10%.
Can I use DMSO as an additive in seal wash?
No. DMSO as an additive in seal wash should be avoided.
When normal-phase solvents are used, IPA or MeOH is recommended for the seal wash solvent.
Can I use DMSO as an additive in needle wash?
Limit DMSO to less than 10%. - wkb196218
ADDITIONAL INFORMATION
This article will be updated as more information becomes available.
DMSO should not be left in a system that will not be used for more than two days.