Liquid on top of FTN vials, no peaks in chromatogram - WKB208892

Article number: 208892

SYMPTOMS

  • Drops of liquid on top of and around vials in sample compartment
  • No peaks in chromatogram

ENVIRONMENT

  • H-Class FTN
  • I-Class FTN

CAUSE

The injection valve pod is not secured tightly by the grub screw. Flow is not being diverted from the needle when the instrument goes to sample from vials.

Needle leaking around barrel

FIX or WORKAROUND

Fit the injection valve cartridge correctly.

Replace needle

ADDITIONAL INFORMATION