2545 Binary Gradient Module shows low pressure and fails leak tests - WKB208845

Article number: 208845

SYMPTOMS

  • Pressure does not build when pump is running
  • Leak test failure

ENVIRONMENT

  • 2525/2545 Binary Gradient Module

CAUSE

Pump is not primed correctly.

FIX or WORKAROUND

  1. Prime the pump using the button on the front panel.
  2. Elevate solvents above the pump if possible to aid priming.
  3. Ensure that liquid is present in the eluent inlet lines at the pump and that no bubbles are visible.

ADDITIONAL INFORMATION