Overpressure in flow cell - WKB207670
Article number: 207670
SYMPTOMS
- High/over pressure in flow cell
ENVIRONMENT
- 2996
- 2489
CAUSE
Blockage in flow cell
FIX or WORKAROUND
- Sonicate the entire cell in a beaker of warm water, and then backflush with the cell out of the instrument.
- Dry fully before refitting to instrument.
- Try backflushing the cell at low flow rate before sonicating.
- If blockage persists, replace the flow cell (WAT057919 - 2996 Analytical cell 10 mm). (WAS081140 - 2487/2489 ANALYTICAL CELL)
ADDITIONAL INFORMATION
id207670,