Alliance 2695 gives shifting retention times - WKB207639
Article number: 207639
SYMPTOMS
- Shifting retention times
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
The sample is not totally dissolved in the mobile phase.
FIX or WORKAROUND
Dissolve the sample in the mobile phase and then filter.
ADDITIONAL INFORMATION
Samples are extracted with 100% methanol, but the mobile phase is 5% ACN in water. Some impurities are not dissolved in the mobile phase,which affects column efficiency.