What is the field Pre-Column Volume in the 2695 instrument method in Empower? - WKB207602
Article number: 207602
ENVIRONMENT
- Empower
- 2695 Separations Module
- Pre-Column Volume

ANSWER
The field Pre-Column Volume tells the 2695 Separations Module to start the gradient and delivers this volume before making an injection. Use this parameter when transferring a method from a system that has a delay volume smaller than that of the 2695 Separations Module. This parameter can also be used to reduce the delay volume of the 2695 Separations Module for narrow or microbore columns, because the sample is held in the sample loop (if desired) until the gradient front reaches the sample loop.