ACQUITY UPLC BSM gives "System Under Pressure" error - WKB207569

Article number: 207569

SYMPTOMS

  • Instrument shows correct flow, but pressure is very low
  • Error occurs within a few minutes of starting flow

ENVIRONMENT

  • ACQUITY UPLC BSM
  • ACQUITY I-CLASS BSM
  • ACQUITY H-CLASS QSM
  • ACQUITY M-CLASS BSM

CAUSE

Solvent bottle is out of mobile phase.

FIX or WORKAROUND

  1. Replace empty bottle(s) with new bottle(s) of fresh mobile phase.
  2. Prime each line for five minutes.

ADDITIONAL INFORMATION