Baseline ripple seen with certain instruments - WKB206105
Article number: 206105
SYMPTOMS
- Wavy baseline with certain instruments
- Until now, no problem appeared as long as the analysis was performed under the same conditions.
- Symptoms appear at specific times during gradient analysis
- Baseline stabilizes when the flow rate is decreased
ENVIRONMENT
- Alliance HPLC System
- Arc HPLC System
- ACQUITY APC
- ACQUITY Arc
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- PATROL UPLC Process Analysis System
- ACQUITY Premier
CAUSE
Deterioration of pump consumables
FIX or WORKAROUND
- Perform a pressure monitoring.
- A slight increase in pressure fluctuations can affect the baseline.
- If the pressure fluctuation is large, perform replacement using the pump PM kit.
ADDITIONAL INFORMATION
Search for the part number of the PM kit from the graphical parts locator.