Baseline ripple seen with certain instruments - WKB206105

Article number: 206105

SYMPTOMS

  • Wavy baseline with certain instruments
  • Until now, no problem appeared as long as the analysis was performed under the same conditions.
  • Symptoms appear at specific times during gradient analysis
  • Baseline stabilizes when the flow rate is decreased

ENVIRONMENT

  • Alliance HPLC System
  • Arc HPLC System
  • ACQUITY APC
  • ACQUITY Arc
  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • PATROL UPLC Process Analysis System
  • ACQUITY Premier

CAUSE

Deterioration of pump consumables

FIX or WORKAROUND

  1. Perform a pressure monitoring.
  2. A slight increase in pressure fluctuations can affect the baseline.
  3. If the pressure fluctuation is large, perform replacement using the pump PM kit.

ADDITIONAL INFORMATION

Search for the part number of the PM kit from the graphical parts locator.

id206105, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, baseline noise, Delta, pressure ripple, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARQSM, UPARSMFTN, UPBSMIVD, UPIBSM, UPIBSM+, UPIBSMIVD, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPLCCOLHTR, UPNBINARY, UPPDAARCB, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMIVD, UPSMPMGR