How to calibrate the needle positions on a 2707 Autosampler - WKB200756
Article number: 200756
OBJECTIVE or GOAL
Calibrate the needle positions on a 2707 Autosampler.
ENVIRONMENT
- 2707 Sample Manager
PROCEDURE
- Go to the Waters Console.
- Select 2707 Autosampler > Maintain.
- Select Calibrate Needle Axes.
- Follow the Calibrate Needle Position wizard instructions.
ADDITIONAL INFORMATION
id200756, 2707, 2707SFC