Ghost peak on an ACQUITY UPLC system - WKB200730

Article number: 200730

SYMPTOMS

  • Ghost peak in a blank injection

ENVIRONMENT

  • ACQUITY Arc
  • ACQUITY Arc Bio
  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS

CAUSE

Contaminated in-line filter cartridge

FIX or WORKAROUND

Replace the in-line filter cartridge.

ADDITIONAL INFORMATION