Ghost peak on an ACQUITY UPLC system - WKB200730
Article number: 200730
SYMPTOMS
- Ghost peak in a blank injection
ENVIRONMENT
- ACQUITY Arc
- ACQUITY Arc Bio
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
CAUSE
Contaminated in-line filter cartridge
FIX or WORKAROUND
Replace the in-line filter cartridge.
ADDITIONAL INFORMATION