ACQUITY UPLC Solvent Manager leaks from top of pump head - WKB199284

Article number: 199284

SYMPTOMS

  • ACQUITY UPLC Solvent Manager leaks from the top of the pump head (outlet of pump head to transducer)
  • Shifting retention times

ENVIRONMENT

  • ACQUITY UPLC Binary Solvent Manager (BSM)
  • ACQUITY UPLC I-Class BSM
  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)

CAUSE

Damaged tubing

FIX or WORKAROUND

Replace the pump head to transducer tubing (part 430002472 for I-Class BSM or QSM Bio, 430001120 for QSM).

ADDITIONAL INFORMATION