Split or tailing peaks observed in chromatogram trace when using ACQUITY UPLC H-Class Column Manager in Position 1 - WKB198471

Article number: 198471

SYMPTOMS

  • Split or tailing peaks observed in chromatogram trace when using ACQUITY UPLC H-Class Column Manager in Position 1  
  • All other column positions show satisfactory chromatography traces

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • Empower

CAUSE

The active preheater is partially blocked.

FIX or WORKAROUND

Replace the active preheater.

ADDITIONAL INFORMATION

 

id198471, APH