Skip to main content
Waters

Split or tailing peaks observed in chromatogram trace when using ACQUITY UPLC H-Class Column Manager in Position 1 - WKB198471

Article number: 198471

SYMPTOMS

  • Split or tailing peaks observed in chromatogram trace when using ACQUITY UPLC H-Class Column Manager in Position 1  
  • All other column positions show satisfactory chromatography traces

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • Empower

CAUSE

The active preheater is partially blocked.

FIX or WORKAROUND

Replace the active preheater.

ADDITIONAL INFORMATION

https://support.waters.com/KB_Inst/C..._Preheater_APH

For list of APH part numbers

Not able to find a solution? Click here to request help.