ACQUITY FTN fails needle seal readiness test, normal operating pressure decreases - WKB197663
Article number: 197663
SYMPTOMS
- FTN fails needle seal readiness test with -2000 psi at 0.5 mL/min
- Needle seal characterization passes
- There is a ~2000-psi drop in normal operating pressure of column
- QSM/BSM passes dynamic leak test
- Low sensitivity may be observed if LC-MS tests are run
ENVIRONMENT
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
A leak at the needle seal
FIX or WORKAROUND
Replace the needle and needle seal (700005215).
ADDITIONAL INFORMATION
Characterize needle seal and perform and pass the needle seal readiness and dynamic leak tests at FTN port 6, including needle and seal.
id197663, DLT, UPARBSMFTN, UPARHSMFTN, UPARPFTN, UPARSMFTN