Compression check reports false passing result - WKB19753

Article number: 19753

SYMPTOMS

  • Compression check passes
  • Final pressure is less than 50 psi
  • Delta decay is within 50 psi

ENVIRONMENT

  • Alliance HPLC system
  • 2695 Separations Module

CAUSE

The injector leaks excessively before decay starts.

FIX or WORKAROUND

Replace Tubing, SS, Syringe to V3.

ADDITIONAL INFORMATION

Contact Waters for a local service representative to replace this part.