Compression check reports false passing result - WKB19753
Article number: 19753
SYMPTOMS
- Compression check passes
- Final pressure is less than 50 psi
- Delta decay is within 50 psi
ENVIRONMENT
- Alliance HPLC system
- 2695 Separations Module
CAUSE
The injector leaks excessively before decay starts.
FIX or WORKAROUND
Replace Tubing, SS, Syringe to V3.
ADDITIONAL INFORMATION
Contact Waters for a local service representative to replace this part.