"No Data Acquired" in Empower with QDa - WKB197086

Article number: 197086

SYMPTOMS

  • Error message in Empower: "No Data Acquired"

clipboard_e8c565d97cb942e01a032f83577166397.png

ENVIRONMENT

  • ACQUITY QDa Mass Detector
  • Empower 3 Personal

CAUSE

Decimal symbol in regional settings

FIX or WORKAROUND

Change the decimal symbol from "," to "."