"Failed to Calibrate system volume" error message on ACQUITY UPLC Sample Manager - WKB195843
Article number: 195843
SYMPTOMS
- Error message "Failed to Calibrate system volume" when characterizing the needle and loop
- 10-µL loop, 30-µL needle, 50-µL sample syringe
- VDD calibration and all other leak tests (sample syringe, wash syringe, needle seal) pass
ENVIRONMENT
- ACQUITY UPLC Sample Manager
CAUSE
Incorrect sample loop and/or sample syringe installed. The 50-µL sample syringe volume is to low for characterizing needle and loop.
FIX or WORKAROUND
ADDITIONAL INFORMATION
id195843, UPBSMIVD, UPIBSM, UPIBSM+, UPIBSMIVD, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPSMIVD, UPSMPMGR, UPSMPMGR