Article number: 195835

SYMPTOMS

  • Shifting retention times
  • Wide peak shape
  • Dynamic leak test failure 

ENVIRONMENT

  • ACQUITY UPLC Binary Solvent Manager (BSM)
  • ACQUITY UPLC I-Class BSM
  • ACQUITY UPC2 BSM
  • ACQUITY UPLC H-Class BSM
  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)

CAUSE

Faulty check valve

FIX or WORKAROUND

Replace the check valve on the failing pump head.

ADDITIONAL INFORMATION