Shifting retention time with ACQUITY UPLC solvent managers - WKB195835

Article number: 195835

SYMPTOMS

  • Shifting retention times
  • Wide peak shape
  • Dynamic leak test failure 

ENVIRONMENT

  • ACQUITY UPLC Binary Solvent Manager (BSM)
  • ACQUITY UPLC I-Class BSM
  • ACQUITY UPC2 BSM
  • ACQUITY UPLC H-Class BSM
  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)

CAUSE

Faulty check valve

FIX or WORKAROUND

Replace the check valve on the failing pump head.

ADDITIONAL INFORMATION

 

id195835, A-30SDM, checkvalve, DLT, UPAQSM, UPBINARY, UPBSM+, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL, UPIBSM, UPIBSM, UPIBSM+, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPBSM, UPPQSM, UPQSM, UPQSM, UPQSM+, UPQSM+, UPQSMBIO, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO