"Degasser increased above limit (x.x psi)" error message - WKB195025
Article number: 195025
SYMPTOMS
- "Degasser increased above limit (x.x psi)"
ENVIRONMENT
- ACQUITY UPLC I-Class PLUS
- ACQUITY H-Class
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Air in degasser chamber or lines
FIX or WORKAROUND
- Re-prime all lines in the solvent manager and the sample manager with acetonitrile or methanol for at least five minutes.
- Return to operational solvents.