High pressure on Alliance during wet prime - WKB194850
Article number: 194850
SYMPTOMS
- System pressure is greater than 600 psi during wet prime; the prime flow rate is 7.5 mL/min
- Salt, forming from precipitation, is seen around the inlet and outlet of the Inline Filter Assembly
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
The Filter Insert is clogged by salt from precipitation, resulting in a higher system pressure during system flow or wet prime
FIX or WORKAROUND
Clean, or replace with either the Inline Filter Assembly or the Filter Insert.
ADDITIONAL INFORMATION
WAT035190 - Inline Filter Assembly
WAT088084 - Filter Insert