Leak at the degasser inlet fitting on an Alliance - WKB19376
Article number: 19376
SYMPTOMS
- Leak at the inlet fitting of the degasser chamber
ENVIRONMENT
- 2695 HPLC
- Alliance separations module
CAUSE
Worn ferrule
FIX or WORKAROUND
- Tighten the fitting.
- If that does not work, replace the ferrule (WAT037373 - Tefzel Ferrule 5/pk).