"Pump H/W overpressure" error with ACQUITY UPLC BSM - WKB193719
Article number: 193719
SYMPTOMS
- "Pump H/W overpressure error" when starting flow
- Removing column, back pressure is as expected for flow rate
- Installing a new column lowers back pressure and error does not appear, but back pressure is still very high
- Upon injection, overpressure error
- Inject valve and needle were replaced
- Solvents are old
ENVIRONMENT
- ACQUITY UPLC BSM
CAUSE
Old solvents with buffers are causing back pressure issue.
FIX or WORKAROUND
Replace old solvents with fresh solvents.
ADDITIONAL INFORMATION
id193719, UPBINARY