Unable to connect to my system in Breeze - WKB1928

Article number: 1928

SYMPTOMS

  • Unable to connect to the system after a reboot

ENVIRONMENT

  • Breeze 2

CAUSE

Reboot of unit in incorrect order

FIX or WORKAROUND

  1. Power-down everything.
  2. Power-on any IEEE components and allow them to complete initialization.
  3. Power-on the computer, but do not open the software.
  4. Power-on any Ethernet components and allow them to complete initialization.
  5. Open software.

ADDITIONAL INFORMATION