ACQUITY 30cm Column Heater/Cooler is not working or communicating - WKB18920
Article number: 18920
SYMPTOMS
- No communication
ENVIRONMENT
- 30cm Column Heater/Cooler
- ARC system
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC I-Class
CAUSE
Column heater was off at sample manager initialization.
FIX or WORKAROUND
Switch on the column H/C first, before the sample manager.