Skip to main content
Waters

Ghost peaks are present when running blanks with a 717 Autosampler - WKB17984

Article number: 17984

SYMPTOMS

  • Ghost peaks are present in blank injections with a 717 Autosampler
  • The analyte dissolves in acetonitrile

ENVIRONMENT

  • 717 Autosampler

CAUSE

There is contamination in the needle of the 717 Autosampler. 

FIX or WORKAROUND

  1. Fill a vial with 100% acetonitrile.
  2. Perform 10 replicate injections to clean out the needle.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.