"Seal not calibrated" error message on ACQUITY UPLC H- or I-Class SM-FTN - WKB17044
Article number: 17044
SYMPTOMS
- "Seal not calibrated" error message
ENVIRONMENT
- ACQUITY UPLC I-Class SM-FTN
- ACQUITY UPLC H-Class SM-FTN
- MassLynx
- Empower
CAUSE
The needle seal calibration value was lost.
FIX or WORKAROUND
In the ACQUITY console:
- Select the SM-FTN in the left-hand pane.
- Select Maintain > Characterize Needle Seal.