e-SAT/IN is not communicating with Empower - WKB1703
Article number: 1703
SYMPTOMS
- "Instrument failure" for the e-SAT/IN
- Green light on the e-SAT/IN is blinking
- In Empower Node properties, on the Instruments tab, the e-SAT/IN says "No" in the "OK?" column
- All other Ethernet instruments also say No in that column
- IEEE instruments say Yes in that column
- e-SAT/IN IP address does not come back in the DHCP when it is restarted
- All Ethernet instruments are connected into an Ethernet switch
ENVIRONMENT
- e-SAT/IN Module
- Empower
CAUSE
- Communication issue with the Ethernet switch
FIX or WORKAROUND
- Reset the Ethernet switch.