Can I control my e2695 over IEEE? - WKB1663

Article number: 1663

ENVIRONMENT

  • Alliance e2695
  • Empower
  • MassLynx

ANSWER

  • Instruments produced before 2025 have IEEE boards and can be controlled using GPIB / HPIB / IEEE-488 connections. 
  • Alliance e2695 instruments produced in 2025 and later do not have an IEEE board so can only be controlled by ethernet
  • Even if you meet the minimum requirements, having other instruments already running over IEEE would be another reason to use it. 
  • Communication over Ethernet is supported with versions of Empower greater than Empower 2 Feature Release 5.
  • MassLynx software cannot use ethernet to control Alliance e2695 instruments so cannot control newer models (with no IEEE board).
  • Cut over serial numbers for eCPUs without the IEEE port are: F25SM7165A (186269505) & E25SM4154A (186269506)

ADDITIONAL INFORMATION

In this context, "IEEE" actually means IEEE-488, also known as GPIB or HPIB.