"Unknown peak" and "Missing peak" when processing in Empower - WKB16191
Article number: 16191
SYMPTOMS
- "Unknown peak" and "Missing peak" for just one peak in review window when processing an injection
- Must constantly update processing method to update retention time
ENVIRONMENT
- Empower 3 Feature Release 2
CAUSE
RT window (set to 2%) is too narrow.
FIX or WORKAROUND
- Edit the processing method and increase the RT window to 3% or 5%, which is the default.
- Save the processing method.
- Reprocess.
ADDITIONAL INFORMATION
id16191, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP