ACQUITY PDA sensitivity is much lower than expected - WKB15649
Article number: 15649
SYMPTOMS
- Peaks have become very small for standards
- 500 lamp hours
ENVIRONMENT
- ACQUITY UPLC System with PDA Detector
CAUSE
Dirty flow cell.
FIX or WORKAROUND
Clean system using the ACQUITY UPLC Cleaning Protocol (attached below)
