ACQUITY Column Manager does not power-on - WKB13925
Article number: 13925
SYMPTOMS
- "Please Reset Instrument" message appears in console for ACQUITY Column Manager
- Column Manager does not turn back on
- Power supply is new
- Plugs and outlets are all functional
ENVIRONMENT
- ACQUITY Column Manager (CM)
CAUSE
The fuses in the plug receptacle are blown.
ADDITIONAL INFORMATION