Microsoft .NET error when accessing ACQUITY console: Object reference not set to an instance of an object - WKB13574

Article number: 13574

SYMPTOMS

  • Microsoft .NET error when accessing the ACQUITY console

clipboard_eea2a0f437e511598907d6a0a476ec159.png

  • Instrument components are ok? No in node properties

ENVIRONMENT

  • Empower
  • Client server environment
  • ACQUITY H-Class

CAUSE

Corrupt ICS Driver

FIX or WORKAROUND

  1. Run the driver pack repair function on the LAC/E (C: Drive > Empower > Instruments >Utilities > DM > DM.exe).
  2. Reboot the LAC/E.
  3. Open the ACQUITY console to confirm that the error is gone.
  4. Run as normal.