ACQUITY BSM/ QSM unable to compress during leak test - WKB12857

Article number: 12857

SYMPTOMS

  • BSM compresses to about 9,000-10,000 psi during dynamic leak test, and then terminates with error: "unable to compress" or "unable to maintain pressure" on primary pump
  • Accumulator pump passes leak test
  • The pump is airtight between the check valve on accumulator pump and the vent valve

ENVIRONMENT

  • ACQUITY UPLC BSM
  • ACQUITY H-Class QSM
  • ACQUITY I-Class BSM
  • ACQUITY Premier
  • Acquity ARC

CAUSE

Traces of rust and scratches on the head support plate or the seal wash housing cause a leak between them.

FIX or WORKAROUND

Replace the seal wash housing (700002597 for classic UPLC; 700009194 for I-class,  700012760 for Acquity ARC) and the head support plate (700002601).

ADDITIONAL INFORMATION