ACQUITY SM-FTN needle leaks - WKB125020

Article number: 125020

SYMPTOMS

  • Lower than expected system pressure
  • Higher pressure when injection valve is in Load position
  • Drops of solvent in sample manager
  • Fails needle seal readiness test with negative value

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY Arc Sample Manager FTN-R (included ARC HPLC)
  • ACQUITY Premier System

CAUSE

  • Solvent leaks through break in needle.

FIX or WORKAROUND

  • Replace the needle 

ADDITIONAL INFORMATION

The part numbers for the SM-FTN and the SM-FTN PLUS needle assemblies are:

  • 700005215 Needle, 15 µL, SS w/ Guide (H-Class and ARC)
  • 700011658 Kit, Textured Needle FTN-15µL MP35N w/guide and seat (H-Class Bio)
  • 700005421 Kit, Needle FTN-15µL MP35N w/guide and seat (H-Class Bio)
  • 700011713Kit, Textured Needle FTN-30µL MP35N w/guide and seat (H-Class Bio)
  • 700011628 Kit, Textured Needle FTN 15µL SS w/guide and seat (H-Class and I-Class PLUS)
  • 700011698 Kit, Textured Needle FTN 30µL SS w/guide and seat (H-Class and I-Class PLUS)
  • 700008977 Assy, Needle FTN 15µL w/guide and seat (I-Class)
  • 700005279 Needle Assembly, 30 µL, FTN w/guide and seat (H-Class, I-Class and ARC)
  • 700012820 Sample Needle Kit, 15-uL, HPS, Textured (Premier standard)
  • 700012821 Sample Needle Kit, 15-uL, HPS (Premier)
  • 700012822 Sample Needle w/Guide and Seat, 30-uL, HPS, Textured (Premier)
  • 700009578 Sample Needle, Arc Bio FTN -30uL, MP35N