Skip to main content
Waters

M-Class nano chromatography poor - WKB124229

Article number: 124229

SYMPTOMS

  • Chromatography poor
  • MS signal height lowered
  • Peaks broad and appear later than expected in run

ENVIRONMENT

  • ACQUITY UPLC M-Class
  • nanoACQUITY UPLC

CAUSE

Autozero incorrect, showing negative flow during set pressure diagnostic

FIX or WORKAROUND

Run autozero on affected pumps.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.